PULSE 6 SERVICE MANUAL
PAGE 7.15
SEPT 2011SUNRISE MEDICAL
Section 7
Tilt Actuator and Micro-Switch Removal
The following procedures describe the removal of
the tilt actuator and/or the micro-switch assembly
used for the creep speed. Read the full set of
directions prior to performing the task. Some
steps may be eliminated if both items do not need
to be replaced. Reverse the procedures to install
the tilt actuator and/or micro-switch and retest as
necessary.
1. Disconnect the 6-pin connector "A" located on
the rear of the tilt system (fi g. 7.38).
2. Remove the cover over the 6 pin connector
on the rear of the tilt system using a 3 mm Allen
wrench (fi gure 7.15.2).
3. Locate the micro-switch assembly and remove
using a 3 mm Allen wrench (fi g. 7.15.3).
Note: this step may be eliminated if the micro-
switch is to be disconnected and left in place.
Removal Procedures (cont.)
fig. 7.15.1
fig. 7.15.2
fig. 7.15.3
A