PULSE 6 SERVICE MANUAL
PAGE 3.5
SEPT 2011SUNRISE MEDICAL
Section 3
VR2 Controller Diagnostics Codes (cont.)
Nine Bars - Solenoid Brake Trip
Battery Fault
Check that the batteries are fully charged, the
correct voltage and in good condition. Take a
voltage reading from pin 1 and pin 2 of the charger
port of the VR2 controller, see (figure 3.5.2) If the
meter reads more than 30 volts, then check the
charger. Otherwise, replace your controller.
Ten Bars - High Battery Voltage
An excessive voltage has been applied to the
control system. This is usually caused by a
poor battery connection. Check the battery
connections.
Fig .3.5.2
2
1
3
The parking brakes have a bad connection. Check
the parking break and motor connections. Make
sure the control system connections are secure.
Measure the two small contacts on the four-pin
motor connector (figure 3.5.1). If both motor
connectors read approximately 60 ohms, (for
chairs prior to PLS-101619 only) then replace
the controller. Otherwise replace the motor that
does not read approximately 60 ohms.
For chairs after PLS-101619 or with serial
number prefix PLS6A, or PLS6B, it should
read 13.3 ohms. If both motor connections
read 13.3 ohms, then replace the controller.
Otherwise replace the motor that does not read
13.3 Ohms.
Fig 3.5.1
Figure 3.5.2 shows Motor in unlocked (free-
wheel) 9bar state.
fig. 3.5.2