PULSE 6 SERVICE MANUAL
PAGE 3.3
SEPT 2011SUNRISE MEDICAL
Section 3
VR2 Controller Diagnostics Codes (cont.)
Otherwise, check the brushes on the Right motor
(figure 3.3.2). Using a Flat head screwdriver
remove the brush cap to access the brush.
Ensure that they are not excessively worn, see
curvature on brush (figure 3.2.1) Replace as
required.
Fig 3.3.1
Check that the batteries are fully charged and
in good condition; and check all cables and
connections. Check the connections to the right
motor, look for a loose or damaged connector.
If the reading is short (resistance is less than
10 K ohms) on any of the readings, proceed to
check the 4-pin motor connector. Measure the
resistance from the bottom contact of the red
thick wire on the 4-pin right motor connector to
each of the top contacts of the connectors see
(figure 3.3.3). Measure the resistance from the
bottom contact of the black thick wire on the 4-pin
right motor connector to each the top contacts of
the connector (below right). If all of the readings
are open, then replace the controller. If any of the
readings are short, then replace the right motor.
Five Bars - Right Motor Wiring Trip
Fig 3.3.3
Test 1
Test 2
Test 3
Test 4
Fig 3.3.2